Intracoronary Optical Coherence Tomography and Histology of Overlapping Everolimus-Eluting Bioresorbable Vascular Scaffolds in a Porcine Coronary Artery Model — Vasim Farooq (2013) | RDL Network
Intracoronary Optical Coherence Tomography and Histology of Overlapping Everolimus-Eluting Bioresorbable Vascular Scaffolds in a Porcine Coronary Artery Model
Objectives
This study sought to assess the vascular response of overlapping Absorb stents compared with overlapping newer-generation everolimus-eluting metallic platform stents (Xience V [XV]) in a porcine coronary artery model.
Background
The everolimus-eluting bioresorbable vascular scaffold (Absorb) is a novel approach to treating coronary lesions. A persistent inflammatory response, fibrin deposition, and delayed endothelialization have been reported with overlapping first-generation drug-eluting stents.
Methods
Forty-one overlapping Absorb and overlapping Xience V (XV) devices (3.0 × 12 mm) were implanted in the main coronary arteries of 17 nonatherosclerotic pigs with 10% overstretch. Implanted coronary arteries were evaluated by optical coherence tomography (OCT) at 28 days (Absorb n = 11, XV n = 7) and 90 days (Absorb n = 11, XV n = 8), with immediate histological evaluation following euthanasia at the same time points. One animal from each time point was evaluated with scanning electron microscopy alone. A total of 1,407 cross sections were analyzed by OCT and 148 cross sections analyzed histologically.
Results
At 28 days in the overlap, OCT analyses indicated 80.1% of Absorb struts and 99.4% of XV struts to be covered (p < 0.0001), corresponding to histological observations of struts with cellular coverage of 75.4% and 99.6%, respectively (p < 0.001). Uncovered struts were almost exclusively related to the presence of “stacked” Absorb struts, that is, with a direct overlay configuration. At 90 days, overlapping Absorb and overlapping XV struts demonstrated >99% strut coverage by OCT and histology, with no evidence of a significant inflammatory process, and comparable % volume obstructions.
Conclusions
In porcine coronary arteries implanted with overlapping Absorb or overlapping XV struts, strut coverage is delayed at 28 days in overlapping Absorb, dependent on the overlay configuration of the thicker Absorb struts. At 90 days, both overlapping Absorb and overlapping XV have comparable strut coverage. The implications of increased strut thickness may have important clinical and design considerations for bioresorbable platforms.
Takayuki Okamura, Yoshinobu Onuma, Héctor M. García‐García, Evelyn Regar, Joanna J. Wykrzykowska, Jacques Koolen, Leif Thuesen, Stefan Windecker, Robert Whitbourn, Dougal R. McClean, John A. Ormiston, Patrick W. Serruys
Josep Gómez‐Lara, Salvatore Brugaletta, Vasim Farooq, Yoshinobu Onuma, Roberto Diletti, Stephan Windecker, Leif Thuesen, Dougal McClean, Jacques Koolen, Robert Whitbourn, Dariusz Dudek, Pieter C. Smits, Bernard Chevalier, Evelyn Regar, Susan Veldhof, Richard Rapoza, John A. Ormiston, Héctor M. García‐García, Patrick W. Serruys
Josep Gómez‐Lara, Salvatore Brugaletta, Roberto Diletti, Scot Garg, Y. Onuma, Bill D. Gogas, Robert J. Van Geuns, Cécile Dorange, Susan Veldhof, Richard Rapoza, Robert Whitbourn, Stephan Windecker, Héctor M. García‐García, Evelyn Regar, Patrick W. Serruys
Takashi Muramatsu, Yoshinobu Onuma, Robert‐Jan van Geuns, Bernard Chevalier, Tejas Patel, Ashok Seth, Roberto Diletti, Héctor M. García‐García, Cécile Dorange, Susan Veldhof, Wai-Fung Cheong, Yukio Ozaki, Robert Whitbourn, Antonio L. Bartorelli, Gregg W. Stone, Alexandre Abizaid, Patrick W. Serruys
Discussion(0)
No comments yet. Be the first to comment.