Usefulness of 64-Slice Multislice Computed Tomography Coronary Angiography to Assess In-Stent Restenosis
Journal of the American College of Cardiology 49(22): 2204-2210
Article 2007 English
Authors
FC
Filippo Cademartiri
JS
Joanne D. Schuijf
FP
Francesca Pugliese
Abstract
1 min read
Objectives
This study sought to evaluate the diagnostic accuracy of 64-slice multislice computed tomography (MSCT) coronary angiography in the follow-up of patients with previous coronary stent implantation.
Background
Recent investigations have shown increased image quality and diagnostic accuracy for noninvasive coronary angiography with 64-slice MSCT as compared with previous-generation MSCT scanners, but data on the evaluation of coronary stents are scarce.
Methods
In 182 patients (152 [84%] male, ages 58 ± 11 years) with previous stent (≥2.5 mm diameter) implantation (n = 192), 64-slice MSCT angiography using either a Sensation 64 (Siemens, Forchheim, Germany) or Aquilion 64 (Toshiba, Otawara, Japan) was performed. At each center, coronary stents were evaluated by 2 experienced observers and evaluated for the presence of significant (≥50%) in-stent restenosis. Quantitative coronary angiography served as the standard of reference.
Results
A total of 14 (7.3%) stented segments were excluded because of poor image quality. In the interpretable stents, 20 of the 178 (11.2%) evaluated stents were significantly diseased, of which 19 were correctly detected by 64-slice MSCT. Accordingly, sensitivity, specificity, and positive and negative predictive value to identify in-stent restenosis in interpretable stents were 95.0% (95% confidence interval [CI] 85% to 100%), 93.0% (95% CI 90% to 97%), 63.3% (95% CI 46% to 81%), and 99.3% (95% CI 98% to 100%), respectively.
Conclusions
In-stent restenosis can be evaluated with 64-slice MSCT with good diagnostic accuracy. In particular, a high negative predictive value of 99% was observed, indicating that 64-slice MSCT may be most valuable as a noninvasive method of excluding in-stent restenosis.
Carlos A.G. van Mieghem, Filippo Cademartiri, Nico R. Mollet, Patrizia Malagutti, Marco Valgimigli, Willem B. Meijboom, Francesca Pugliese, Eugène McFadden, Jürgen Ligthart, Giuseppe Runza, Nico Bruining, Pieter C. Smits, Evelyn Regar, Willem J. van der Giessen, Georgios Sianos, Ron van Domburg, Peter de Jaegere, Gabriël P. Krestin, Patrick W. Serruys, Pim J. de Feyter
Nico R. Mollet, Filippo Cademartiri, Koen Nieman, Francesco Saia, Pedro A. Lemos, Eugène McFadden, Peter M. T. Pattynama, Patrick W. Serruys, Gabriël P. Krestin, Pim J. de Feyter
Koen Nieman, Benno J. Rensing, Robert‐Jan van Geuns, Arie Munne, Jürgen Ligthart, Peter M. T. Pattynama, Gabriël P. Krestin, Patrick W. Serruys, Pim J. de Feyter
Nico R. Mollet, Filippo Cademartiri, Carlos A.G. van Mieghem, Giuseppe Runza, Eugène McFadden, Timo Baks, Patrick W. Serruys, Gabriël P. Krestin, Pim J. de Feyter
Discussion(0)