The Impact of Post-Procedural Asymmetry, Expansion, and Eccentricity of Bioresorbable Everolimus-Eluting Scaffold and Metallic Everolimus-Eluting Stent on Clinical Outcomes in the ABSORB II Trial — Pannipa Suwannasom (2016) | RDL Network
The Impact of Post-Procedural Asymmetry, Expansion, and Eccentricity of Bioresorbable Everolimus-Eluting Scaffold and Metallic Everolimus-Eluting Stent on Clinical Outcomes in the ABSORB II Trial
КАРДИОЛОГИЯ УЗБЕКИСТАНА 9(12): 1231-1242
Article 2016 English
Authors
PS
Pannipa Suwannasom
YS
Yohei Sotomi
YI
Yuki Ishibashi
Abstract
2 min read
Objectives
The study sought to investigate the relationship between post-procedural asymmetry, expansion, and eccentricity indices of metallic everolimus-eluting stent (EES) and bioresorbable vascular scaffold (BVS) and their respective impact on clinical events at 1-year follow-up.
Background
Mechanical properties of a fully BVS are inherently different from those of permanent metallic stent.
Methods
The ABSORB II (A bioresorbable everolimus-eluting scaffold versus a metallic everolimus-eluting stent for ischaemic heart disease caused by de-novo native coronary artery lesions) trial compared the BVS and metallic EES in the treatment of a de novo coronary artery stenosis. Protocol-mandated intravascular ultrasound imaging was performed pre- and post-procedure in 470 patients (162 metallic EES and 308 BVS). Asymmetry index (AI) was calculated per lesion as: (1 − minimum scaffold/stent diameter/maximum scaffold/stent diameter). Expansion index and optimal scaffold/stent expansion followed the definition of the MUSIC (Multicenter Ultrasound Stenting in Coronaries) study. Eccentricity index (EI) was calculated as the ratio of minimum and maximum scaffold/stent diameter per cross section. The incidence of device-oriented composite endpoint (DoCE) was collected.
Results
Post-procedure, the metallic EES group was more symmetric and concentric than the BVS group. Only 8.0% of the BVS arm and 20.0% of the metallic EES arm achieved optimal scaffold/stent expansion (p < 0.001). At 1 year, there was no difference in the DoCE between both devices (BVS 5.2% vs. EES 3.1%; p = 0.29). Post-procedural devices asymmetry and eccentricity were related to higher event rates while there was no relevance to the expansion status. Subsequent multivariate analysis identified that post-procedural AI >0.30 is an independent predictor of DoCE (hazard ratio: 3.43; 95% confidence interval: 1.08 to 10.92; p = 0.037).
Conclusions
BVS implantation is more frequently associated with post-procedural asymmetric and eccentric morphology compared to metallic EES. Post-procedural devices asymmetry were independently associated with DoCE following percutaneous coronary intervention. However, this approach should be viewed as hypothesis generating due to low event rates. (ABSORB II Randomized Controlled Trial [ABSORB II]; NCT01425281)
Takashi Muramatsu, Yoshinobu Onuma, Héctor M. García‐García, Vasim Farooq, Christos V. Bourantas, Marie-angèle Morel, Xiaolin Li, Susan Veldhof, Antonio L. Bartorelli, Robert Whitbourn, Alexandre Abizaid, Patrick W. Serruys
Takashi Muramatsu, Yoshinobu Onuma, Robert‐Jan van Geuns, Bernard Chevalier, Tejas Patel, Ashok Seth, Roberto Diletti, Héctor M. García‐García, Cécile Dorange, Susan Veldhof, Wai-Fung Cheong, Yukio Ozaki, Robert Whitbourn, Antonio L. Bartorelli, Gregg W. Stone, Alexandre Abizaid, Patrick W. Serruys
Yoshinobu Onuma, Patrick W. Serruys, J.A. Fernández Gómez, Bernard De Bruyne, Dariusz Dudek, Leif Thuesen, Peter Smits, Bernard Chevalier, Dougal McClean, Jacques Koolen, Stephan Windecker, Robert Whitbourn, Ian T. Meredith, Héctor M. García‐García, John A. Ormiston
Kyohei Yamaji, Salvatore Brugaletta, Manel Sabaté, Andrés Íñiguez, Lisette Okkels Jensen, Ángel Cequier, Sjoerd H. Hofma, Evald Høj Christiansen, Maarten J. Suttorp, Gerrit Anne van Es, Yohei Sotomi, Yoshinobu Onuma, Patrick W. Serruys, Stephan Windecker, Lorenz Räber
Salvatore Brugaletta, Tommaso Gori, Adrian F. Low, Petr Toušek, Eduardo Pinar, Josep Gómez‐Lara, Giancarla Scalone, Eberhard Schulz, Mark Y. Chan, Viktor Kočka, José Hurtado, Juan Antoni Gomez-Hospital, Thomas Münzel, Chi‐Hang Lee, Ángel Cequier, Mariano Valdés, Petr Widimský, Patrick W. Serruys, Manel Sabaté
Discussion(0)
No comments yet. Be the first to comment.