Reproducibility of exhaled nitric oxide measurements in overweight and obese adults
BMC Research Notes 7(1)
Article 2014 English
Authors
WT
Willemien Thijs
RM
Renée de Mutsert
SC
Saskia le Cessie
Abstract
1 min read
Exhaled nitric oxide is a noninvasive measure of airway inflammation that can be detected by a handheld device. Obesity may influence the reproducibility of exhaled nitric oxide measurements, by - for instance – decreased expiratory reserve volume. We analyzed triple exhaled nitric oxide measurements from 553 participants (aged 45 to 65 years with a body mass index ≥27 kg/m2) of the Netherlands Epidemiology of Obesity Study. The interclass correlation coefficient (single measurement reliability) was 0.965 (95% CI: 0.960, 0.970). We conclude that for assessment of exhaled nitric oxide in large cohorts of overweight and obese adults a single measurement suffices.