Flexor Tendon Repair Using a Stainless Steel External Splint
Journal of Hand Surgery (European Volume) 24(6): 654-657
Article 1999 English
Authors
LG
Leonard Gordon
FD
F. A. DYSARZ
KV
K. T. Venkateswara
Abstract
1 min read
A stainless steel external tendon splint was used in repair of cadaver tendons and compared with standard tendon repairs with suture. The splint was combined with a Kessler repair and tested against the Kessler, Becker, and Savage repairs in fresh human cadaver flexor digitorum profundus tendons. Biomechanical testing was done on a tensile testing machine, and load-displacement curves were generated. The repairs using the external tendon splint demonstrated a range of improvement of 32 to 146% in mean maximal tensile strength and a 20 to 185% improvement of mean ultimate tensile strength compared with all other repairs. The external tendon splint is relatively easy to apply to a tendon. The repair is strengthened and becomes capable of withstanding early active range of motion exercises. In vivo testing will be needed to assess the potential clinical usefulness of such a device.