Background Individualized hemodynamic management during surgery relies on accurate titration of vasopressors and fluids. In this context, computer systems have been developed to assist anesthesia providers in delivering these interventions. This study tested the hypothesis that computer-assisted individualized hemodynamic management could reduce intraoperative hypotension in patients undergoing intermediate- to high-risk surgery. Methods This single-center, parallel, two-arm, prospective randomized controlled single blinded superiority study included 38 patients undergoing abdominal or orthopedic surgery. All included patients had a radial arterial catheter inserted after anesthesia induction and connected to an uncalibrated pulse contour monitoring device. In the manually adjusted goal-directed therapy group (N = 19), the individualized hemodynamic management consisted of manual titration of norepinephrine infusion to maintain mean arterial pressure within 10% of the patient’s baseline value, and mini-fluid challenges to maximize the stroke volume index. In the computer-assisted group (N = 19), the same approach was applied using a closed-loop system for norepinephrine adjustments and a decision-support system for the infusion of mini-fluid challenges (100 ml). The primary outcome was intraoperative hypotension defined as the percentage of intraoperative case time patients spent with a mean arterial pressure of less than 90% of the patient’s baseline value, measured during the preoperative screening. Secondary outcome was the incidence of minor postoperative complications. Results All patients were included in the analysis. Intraoperative hypotension was 1.2% [0.4 to 2.0%] (median [25th to 75th] percentiles) in the computer-assisted group compared to 21.5% [14.5 to 31.8%] in the manually adjusted goal-directed therapy group (difference, −21.1 [95% CI, −15.9 to −27.6%]; P < 0.001). The incidence of minor postoperative complications was not different between groups (42 vs. 58%; P = 0.330). Mean stroke volume index and cardiac index were both significantly higher in the computer-assisted group than in the manually adjusted goal-directed therapy group ( P < 0.001). Conclusions In patients having intermediate- to high-risk surgery, computer-assisted individualized hemodynamic management significantly reduces intraoperative hypotension compared to a manually controlled goal-directed approach. Editor’s Perspective What We Already Know about This Topic What This Article Tells Us That Is New
Alexandre Joosten, Shalini Raj Lawrence, Alexandra Colesnicenco, Sean Coeckelenbergh, Jean Louis Vincent, Philippe Van der Linden, Maxime Cannesson, Joseph Rinehart
Alexandre Joosten, Valério Lucidi, Brigitte Ickx, Luc Van Obbergh, Desislava Germanova, Antoine Berna, Brenton Alexander, Olivier Desebbe, Francois-martin Carrier, Daniel Cherqui, René Adam, Jacques Duranteau, Bernd Saugel, Jean Louis Vincent, Joseph Rinehart, Philippe Van der Linden
Sean Coeckelenbergh, Joseph Rinehart, Olivier Desebbe, Nicolas Rogoz, Amira Dagachi Mastouri, Bryan Maghen, Maxime Cannesson, Jean Louis Vincent, Jacques Duranteau, Alexandre Joosten
Olivier Desebbe, Joseph Rinehart, Philippe Van der Linden, Maxime Cannesson, Bertrand Delannoy, Marc Vigneron, Alain Curtil, Etienne Hautin, Jean Louis Vincent, Jacques Duranteau, Alexandre Joosten
E Osawa, A Rhodes, Jean Louis Vincent, Juliano Pinheiro de Almeida, J Fukushima, Brunna Pileggi, C Park, Leandro Teixeira Câmara, José Otávio Costa Auler Júnior, Raquel Pei Chen Chan, M Piccioni, M. Isabel P. de Lima, F Galas, Ludhmila Abrahão Hajjar
Discussion(0)
No comments yet. Be the first to comment.