Angular stable lateral plating is a valid alternative to conventional plate fixation in the proximal phalanx. A biomechanical study — Rukmanikanthan Shanmugam (2015) | RDL Network
Angular stable lateral plating is a valid alternative to conventional plate fixation in the proximal phalanx. A biomechanical study
Clinical Biomechanics 30(5): 405-410
Article 2015 English
Authors
RS
Rukmanikanthan Shanmugam
ME
Manuela Ernst
KS
Karl Stoffel
Abstract
1 min read
Background
Dorsal plating is commonly used in proximal phalanx fractures but it bears the risk of interfering with the extensor apparatus. In this study, dorsal and lateral plating fixation methods are compared to assess biomechanical differences using conventional 1.5mm non-locking plates and novel 1.3mm lateral locking plates.
Methods
Twenty-four fresh frozen human cadaveric proximal phalanges were equally divided into four groups. An osteotomy was set at the proximal metaphyseal–diaphyseal junction and fixed with either dorsal (group A) or lateral (group B) plating using a 1.5mm non-locking plate, or lateral plating with a novel 1.3mm locking plate with bicortical (group C) or unicortical (group D) screws. The specimens were loaded in axial, dorsovolar and mediolateral direction to assess fixation stiffness followed by a cyclic destructive test in dorsovolar loading direction.
Findings
Axial stiffness was highest in group D (mean 321.02, SEM 21.47N/mm) with a significant difference between groups D and B (P
=0.033). Locking plates (groups C and D) were stiffer than non-locking plates under mediolateral loading (P
=0.007), no significant differences were noted under dorsovolar loading. Furthermore, no significant differences were observed under cyclic loading to failure between any of the study groups.
Interpretation
No considerable biomechanical advantage of using a conventional 1.5mm dorsal non-locking plate was identified over the novel 1.3mm lateral locking plate in the treatment of proximal phalanx fractures. Since the novel low-profile plate is less disruptive to the extensor mechanism, it should be considered as a valid alternative.
Jiayue Huang, Fan Yang, Juan Luis Gutiérrez‐Chico, Tianxiao Xu, Jigang Wu, Liang Wang, Rui Lv, Yan Lai, Xuebo Liu, Yoshinobu Onuma, Dalin Tang, Patrick W. Serruys, William Wijns, Shengxian Tu
Discussion(0)