Acute Gain in Minimal Lumen Area Following Implantation of Everolimus-Eluting ABSORB Biodegradable Vascular Scaffolds or Xience Metallic Stents — Yohei Sotomi (2016) | RDL Network
Acute Gain in Minimal Lumen Area Following Implantation of Everolimus-Eluting ABSORB Biodegradable Vascular Scaffolds or Xience Metallic Stents
КАРДИОЛОГИЯ УЗБЕКИСТАНА 9(12): 1216-1227
Article 2016 English
Authors
YS
Yohei Sotomi
YI
Yuki Ishibashi
PS
Pannipa Suwannasom
Abstract
2 min read
Objectives
The study compared, by intravascular ultrasound (IVUS), acute gain (AG) at the site of the pre-procedural minimal lumen area (MLA) achieved by either the Absorb (Abbott Vascular, Santa Clara, California) scaffold or the Xience stent and identified the factors contributing to the acute performance of these devices.
Background
It is warranted that the acute performance of Absorb matches that of metallic stents; however, concern exists about acute expansion and lumen gain with the use of Absorb.
Methods
Of a total of 501 patients (546 lesions) in the ABSORB II (ABSORB II Randomized Controlled Trial) randomized trial, 445 patients with 480 lesions were investigated by IVUS pre- and post-procedure. Comparison of MLA pre- and post-procedure was performed at the MLA site by matching pre- and post-procedural IVUS pullbacks.
Results
Lower AG on IVUS (lowest tertile) occurred more frequently in the Absorb arm than in the Xience arm (3.46 mm2 vs. 4.27 mm2, respectively; p < 0.001; risk ratio: 3.04; 95% confidence interval: 1.94 to 4.76). The plaque morphology at the MLA cross-section was not independently associated with IVUS acute gain. The main difference in AG in MLD by angiography was observed at the time of device implantation (Xience vs. Absorb, Δ+1.50 mm vs. Δ+1.23 mm, respectively), whereas the gain from post-dilation was similar between the 2 arms (Δ+0.16 mm vs. Δ+0.16 mm) when patients underwent post-dilation, although expected balloon diameter was smaller in the Absorb arm than in the Xience arm (p = 0.003) during post-dilation.
Conclusions
At the site of the pre-procedural MLA, the increase of the lumen post-procedure was smaller in the Absorb-arm than in the Xience arm. To achieve equivalent AG to Xience, the implantation of Absorb may require more aggressive strategies at implantation, pre- and post-dilation than the technique used in the ABSORB II trial. (ABSORB II Randomized Controlled Trial [ABSORB II]; NCT01425281)
Yuki Katagiri, Patrick W. Serruys, Erhan Tenekecioğlu, Taku Asano, Carlos Collet, Yosuke Miyazaki, Athanasios Katsikis, Jan J. Piek, Joanna J. Wykrzykowska, Bernard Chevalier, Gary S. Mintz, Gregg W. Stone, Yoshinobu Onuma
Yohei Sotomi, Yoshinobu Onuma, Pannipa Suwannasom, Hiroki Tateishi, Erhan Tenekecioğlu, Yaping Zeng, Rafael Cavalcante, Hans Jonker, Jouke Dijkstra, Nicolas Foin, Jaryl Ng Chen Koon, Carlos Collet, Robbert J. de Winter, Joanna J. Wykrzykowska, Gregg W. Stone, Jeffrey J. Popma, Ken Kozuma, Kengo Tanabe, Patrick W. Serruys, Takeshi Kimura
Kyohei Yamaji, Salvatore Brugaletta, Manel Sabaté, Andrés Íñiguez, Lisette Okkels Jensen, Ángel Cequier, Sjoerd H. Hofma, Evald Høj Christiansen, Maarten J. Suttorp, Gerrit Anne van Es, Yohei Sotomi, Yoshinobu Onuma, Patrick W. Serruys, Stephan Windecker, Lorenz Räber
Yao‐Jun Zhang, Christos V. Bourantas, Takashi Muramatsu, Javaid Iqbal, Vasim Farooq, Roberto Diletti, Carlos M. Campos, Yoshinobu Onuma, Héctor M. García‐García, Patrick W. Serruys
Roberto Diletti, Yoshinobu Onuma, Vasim Farooq, Josep Gómez‐Lara, Salvatore Brugaletta, Robert‐Jan van Geuns, Evelyn Regar, Bernard De Bruyne, Dariusz Dudek, Leif Thuesen, Bernard Chevalier, Dougal McClean, Stephan Windecker, Robert Whitbourn, Pieter C. Smits, Jacques Koolen, Ian T. Meredith, Dong Li, Susan Veldhof, Richard Rapoza, Héctor M. García‐García, John A. Ormiston,