A Prospective Randomized Trial Comparing Sirolimus-Coated Balloon With Paclitaxel-Coated Balloon in De Novo Small Vessels — Kai Ninomiya (2023) | RDL Network
Background
There are no data comparing sirolimus-coated balloons (SCBs [MagicTouch, Concept Medical]) to paclitaxel-coated balloons (PCBs [SeQuent Please Neo, B. Braun]) for the treatment of de novo small vessel disease (SVD).
Objectives
This study sought to compare quantitative coronary angiographic outcomes at 6 months after treatment of de novo SVD with a PCB or SCB.
Methods
This prospective, multicenter, noninferiority trial randomized 121 patients (129 SVD lesions) to treatment with an SCB or PCB, with balloon sizing determined using optical coherence tomography. The primary endpoint was noninferiority for the 6-month angiographic net lumen gain.
Results
Angiographic follow-up was completed in 109 (90.1%) patients in the per-protocol analysis. The mean ± SD angiographic net gains were 0.25 ± 0.40 mm with SCBs vs 0.48 ± 0.37 mm with PCBs, resulting in SCBs failing to meet the 0.30 mm criterion for noninferiority (P
noninferiority = 0.173), with an absolute difference of −0.23 mm (95% CI: −0.37 to −0.09) secondary to a smaller late loss (0.00 ± 0.32 mm vs 0.32 ± 0.47 mm; P < 0.001) and more frequent late lumen enlargement (53.7% vs 30.0%; OR: 2.60; 95% CI: 1.22-5.67; P = 0.014) with PCBs. Binary restenosis rates were 32.8% and 12.5% following treatment with SCBs and PCBs, respectively (OR: 3.41; 95% CI: 1.36-9.44; P = 0.012). The mean angiography-derived fractional flow ratio at follow-up was 0.86 ± 0.15 following treatment with SCBs and 0.91 ± 0.09 following PCBs (P = 0.026); a fractional flow ratio ≤0.80 occurred in 13 and 5 vessels after treatment with SCBs and PCBs, respectively.
Conclusions
The SCB MagicTouch failed to demonstrate noninferiority for angiographic net lumen gain at 6 months compared to the PCB SeQuent Please Neo.
Marie-Claude Morice, Antonio Colombo, Bernhard Meier, Patrick W. Serruys, Corrado Tamburino, Giulio Guagliumi, Eduardo Sousa, Hans-Peter Stoll, for the REALITY Trial Investigators