A Comparison of the Conformability of Everolimus-Eluting Bioresorbable Vascular Scaffolds to Metal Platform Coronary Stents — Josep Gómez‐Lara (2010) | RDL Network
A Comparison of the Conformability of Everolimus-Eluting Bioresorbable Vascular Scaffolds to Metal Platform Coronary Stents
КАРДИОЛОГИЯ УЗБЕКИСТАНА 3(11): 1190-1198
Article 2010 English
Authors
JG
Josep Gómez‐Lara
HG
Héctor M. García‐García
YO
Yoshinobu Onuma
Abstract
1 min read
Objectives
The aim of this study was to assess the differences in terms of curvature and angulation of the treated vessel after the deployment of either a metallic stent or a polymeric scaffold device.
Background
Conformability of metallic platform stents (MPS) is the major determinant of geometric changes in coronary arteries caused by the stent deployment. It is not known how bioresorbable polymeric devices perform in this setting.
Methods
This retrospective study compares 102 patients who received an MPS (Multi-link Vision or Xience V, Abbott Vascular, Santa Clara, California) in the SPIRIT FIRST and II trials with 89 patients treated with the Revision 1.1 everolimus-eluting bioresorbable vascular scaffold (BVS) (Abbott Vascular, Santa Clara, California) from cohort B of the ABSORB (A bioabsorbable everolimus-eluting coronary stent system) trial. All patients were treated with a single 3 × 18 mm device. Curvature and angulation were measured with dedicated software by angiography.
Results
Both the MPS and BVS groups had significant changes in relative region curvature (MPS vs. BVS: 28.7% vs. 7.5%) and angulation (MPS vs. BVS: 25.4% vs. 13.4%) after deployment. The unadjusted comparisons between the 2 groups showed for BVS a nonsignificant trend for less change in region curvature after deployment (MPS vs. BVS: 0.085 cm−1 vs. 0.056 cm−1, p = 0.06) and a significantly lower modification of angulation (MPS vs. BVS 6.4° vs. 4.3°, p = 0.03). By multivariate regression analysis, the independent predictors of changes in curvature and angulation were the pre-treatment region curvature, the pre-treatment region angulation, and the used device.
Conclusions
Bioresorbable vascular scaffolds have better conformability than conventional MPS. The clinical significance of the observed differences will require further investigation.
Josep Gómez‐Lara, Salvatore Brugaletta, Vasim Farooq, Robert‐Jan van Geuns, Bernard De Bruyne, Stephan Windecker, Dougal McClean, Leif Thuesen, Dariusz Dudek, Jacques Koolen, Robert Whitbourn, Pieter C. Smits, Bernard Chevalier, Marie‐Angèle Morel, Cécile Dorange, Susan Veldhof, Richard Rapoza, Héctor M. García‐García, John A. Ormiston, Patrick W. Serruys
Josep Gómez‐Lara, Salvatore Brugaletta, Vasim Farooq, Yoshinobu Onuma, Roberto Diletti, Stephan Windecker, Leif Thuesen, Dougal McClean, Jacques Koolen, Robert Whitbourn, Dariusz Dudek, Pieter C. Smits, Bernard Chevalier, Evelyn Regar, Susan Veldhof, Richard Rapoza, John A. Ormiston, Héctor M. García‐García, Patrick W. Serruys
Takashi Muramatsu, Yoshinobu Onuma, Robert‐Jan van Geuns, Bernard Chevalier, Tejas Patel, Ashok Seth, Roberto Diletti, Héctor M. García‐García, Cécile Dorange, Susan Veldhof, Wai-Fung Cheong, Yukio Ozaki, Robert Whitbourn, Antonio L. Bartorelli, Gregg W. Stone, Alexandre Abizaid, Patrick W. Serruys
Takashi Muramatsu, Yoshinobu Onuma, Héctor M. García‐García, Vasim Farooq, Christos V. Bourantas, Marie-angèle Morel, Xiaolin Li, Susan Veldhof, Antonio L. Bartorelli, Robert Whitbourn, Alexandre Abizaid, Patrick W. Serruys
Salvatore Brugaletta, Tommaso Gori, Adrian F. Low, Petr Toušek, Eduardo Pinar, Josep Gómez‐Lara, Giancarla Scalone, Eberhard Schulz, Mark Y. Chan, Viktor Kočka, José Hurtado, Juan Antoni Gomez-Hospital, Thomas Münzel, Chi‐Hang Lee, Ángel Cequier, Mariano Valdés, Petr Widimský, Patrick W. Serruys, Manel Sabaté
Discussion(0)
No comments yet. Be the first to comment.