A bioabsorbable everolimus-eluting coronary stent system for patients with single de-novo coronary artery lesions (ABSORB): a prospective open-label trial — John A. Ormiston (2008) | RDL Network
A bioabsorbable everolimus-eluting coronary stent system for patients with single de-novo coronary artery lesions (ABSORB): a prospective open-label trial
Background
A fully bioabsorbable drug-eluting coronary stent that scaffolds the vessel wall when needed and then disappears once the acute recoil and constrictive remodelling processes have subsided has theoretical advantages. The bioasorbable everolimus-eluting stent (BVS) has a backbone of poly-L-lactic acid that provides the support and a coating of poly-D,L-lactic acid that contains and controls the release of the antiproliferative agent everolimus. We assessed the feasibility and safety of this BVS stent.
Methods
In this prospective, open-label study we enrolled 30 patients who had either stable, unstable, or silent ischaemia and a single de-novo lesion that was suitable for treatment with a single 3·0×12 mm or 3·0×18 mm stent. Patients were enrolled from four academic hospitals in Auckland, Rotterdam, Krakow, and Skejby. The composite endpoint was cardiac death, myocardial infarction, and ischaemia-driven target lesion revascularisation. Angiographic endpoints were available for 26 patients and intravascular-ultrasound endpoints for 24 patients. Clinical endpoints were assessed in all 30 patients at 6 and 12 months. In a subset of 13 patients, optical coherence tomography was undertaken at baseline and follow-up. Analysis was by intention to treat. This study is registered with ClinicalTrials.gov, number NCT00300131.
Findings
Procedural success was 100% (30/30 patients), and device success 94% (29/31 attempts at implantation of the stent). At 1 year, the rate of major adverse cardiac events was 3·3%, with only one patient having a non-Q wave myocardial infarction and no target lesion revascularisations. No late stent thromboses were recorded. At 6-month follow-up, the angiographic in-stent late loss was 0·44 (0·35) mm and was mainly due to a mild reduction of the stent area (−11·8%) as measured by intravascular ultrasound. The neointimal area was small (0·30 [SD 0·44] mm2), with a minimal area obstruction of 5·5%.
Interpretation
This study shows the feasibility of implantation of the bioabsorbable everolimus-eluting stent, with an acceptable in-stent late loss, minimal intrastent neointimal hyperplasia, and a low stent area obstruction.
Funding
Abbott Vascular.
Patrick W. Serruys, John A. Ormiston, Yoshinobu Onuma, Evelyn Regar, Nieves Gonzalo, Héctor M. García‐García, Koen Nieman, Nico Bruining, Cécile Dorange, Karine Miquel‐Hébert, Susan Veldhof, Mark Webster, Leif Thuesen, Dariusz Dudek
Patrick W. Serruys, P. Ruygrok, Jörg Neuzner, Jan J. Piek, Ashok Seth, Joachim Schofer, Gert Richardt, Marcus Wiemer, Didier Carrié, Leif Thuesen, Els Boone, Karine Miquel-Herbert, Joost Daemen
Jörg Neuzner, Jan J. Piek, Ashok Seth, Joachim Schofer, Gert Richardt, Marcus Wiemer, Didier Carrié, Leif Thuesen, Els Boone, Karine Miquel-Herbert, Joost Daemen, Patrick W. Serruys, P. Ruygrok
Keiichi Tsuchida, Jan J. Piek, Franz‐Josef Neumann, Willem J. van der Giessen, Marcus Wiemer, Andreas M. Zeiher, Eberhard Grube, Jürgen Haase, Leif Thuesen, Christian W. Hamm, Susan Veldhof, Cécile Dorange, Patrick W. Serruys